Features :
Performance
- It is designed to be suitable for the storage of Red cell Concentration at 4 °c
- Our blood bank refrigerator provide highly accurate temperator control and distribution using original techniques developed by Daido Industries INC.
- Maximum storage of DRS-1380 is 600 blood bags.
- Vacuum insulated double glass doors prevent the outer heat and provide easily to see the inside.
Safety Device :
- Our blood bank refrigerator has dual safety devices for temperature control in the chamber
- The alarm buzzer activates when the inside temperature does not keep within the normal range (2.2°c to 5.8 °c) and the comperssor shutdown automatically st 2.0 °c or 12.0 °c for quality of RCC
- DRS-1380 has 2 separated cooling system, even though 1 system has something trouble, but the other system can maintain the inside setting temperature.
Specification | ||
Model | DRS-600 | DRS-1380 |
Certificate no. | 21900BZX00008000 | |
Capacity | 570L | 1380L |
Materials | SUS304 stainless steel | |
Door | Vacuum insulated double glass doors | |
1 door | 2 door | |
Insulator | polyrethane | |
Cooling | Refrigerant gas temperature control , Horizontal laminar blow | |
Heat radiation | Air cooled condenser | |
Refrigerant Gas | HFC404A (CFC Free) | |
Power Source | 1 phase 100V 50/60 Hz Depending on you power source , you might need a transformer and a stabilizer | |
Cooling System | 1 System | Separated 2 system |
Thermostat | Electronic digital 1 system | Electronic digital separated 2 systems |
Control System | 1 System | Separated 2 system |
Alarm | 2 Digital Thermo-alarm / Powewr fail alarm | |
Safety Device | Shut down compressor when over - cooling or over - heating |